公司新聞

有關拉力試驗機常見故障的排除方案

發布時間:2018-11-23 來源:濟南恒思盛大儀器有限公司 瀏覽:

一、拉力試驗機加荷時,指針顫動或時走時停

  1、離合器齒輪磨損:需要修理或更換。

  2、摩擦盤的皮墊圈或彈簧磨損:需更換皮墊圈或彈簧。

  3、操縱手柄移位:調整操縱手柄,使其與牙槽配合好。

  二、更換擺砣時指針不回零

  1、拉力試驗機安裝不水平:用水平儀把試驗機調整水平。

  2、擺錘不垂直:只掛A砣。調整平衡砣,使之垂直。

  三、拉力試驗機擺錘回位太快或太慢

  1、緩沖閥擋位位置放置不當:調整緩沖閥到適當位置。

  2、液壓油黏度過低或過高:液壓油黏度過低時擺錘回落快,黏度高時回落慢.應更換適當黏度的液壓油。

  3、緩沖閥內、油管內或液壓油有臟污:清洗緩沖閥、油管。更換液壓油。http://www.hi-orange.net/